ALSAGLIDE COATING TECHNOLOGY
MULTIKNOT DESIGN
USER FRIENDLY DESIGN FOR EASY LOADING
- Hold cartridge and arrange ‘V’ position with leaving narrow angle in between its two wings. Position of wings shouldn’t be wide open entire procedure.
- Put small amount of ALSAVISC inside cartridge. Not recommended fully fill cartridge
- Fill OVD both cartridge loading bay and groove.Spread OVD within each other. Avoid over fill.
- Pull out IOL gently from holder
- Before placing IOL into cartridge Make sure that holes on haptic or form of haptic indicate “reverse-S” shape
- Place IOL carefully into cartridge loading bay. Ensure that all haptic part of IOL is located below the groove lines. Both optic and Haptic parts should be placed in tunnel.
- Use forceps to apply litte press on haptics to locate IOL in tunnel properly. Always check haptic parts before closing cartridge.Close cartridge by pushing wings until hear ‘click’ sound that indicates proper closing.
- Close cartridge by pushing wings until hear ‘click’ sound that indicates proper closing
- Insert cartridge to path on injector with slight angle. Apply some amount of OVD from cartridge hole if necessary.
- Push plunger slowly and check IOL position and check folding pattern from transparent cartridge lastly before insertion.
IOL MODEL | DIOPTER RANGE | Compatible ALSAJET MODEL | |
MONOFOCAL | ![]() | -20.00 to 30.00D | ALSAJET 2.2 |
![]() | -20.00D to 25.00D -20.00 to 30.00D | ALSAJET 2.0 ALSAJET 2.2 | |
![]() | -20.00D to 28.00D -20.00 to 25.00D -20.00 to 30.00D | ALSAJET 1.8 | |
-20.00 to 30.00D | ALSAJET 2.0 ALSAJET 2.2 | ||
MONOFOCAL TORIC | ![]() | Spherical Power -20.00 to 25.00D Cylinder power +1.00 to +5.00D Spherical Power -20.00D to 28.00D Cylinder Power +1.00D to +5.00D Spherical Power -20.00 to 30D Cylinder power +1.00 to +5.00D | ALSAJET 1.8 ALSAJET 2.0 ALSAJET 2.2 |
MULTIFOCAL | ![]() | 0 to 30.00D | ALSAJET 2.2 |
TRIFOCAL | ![]() | -20.00D to 25.00D -20.00D to 28.00D -20.00D to 30.00D | ALSAJET 1.8 ALSAJET 2.0 ALSAJET 2.2 |